CVE-2026-67558
8.2
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:A/VC:H/VI:L/VA:N/SC:H/SI:L/SA:N
Summary
The Mira Android companion app v4.5.15.4 identifies the paired Mira hormone analyzer by performing a substring match against the BLE advertisement name only, with no cryptographic peripheral authentication, MAC allowlist, or bonded-identity check. An attacker could capture live session token information and inject forged hormone measurements into the victim's cloud record and clinical trend view.
Affected Software
| Vendor | Product | Version Range | Status |
|---|---|---|---|
| Quanovate Tech Inc. (operating as Mira / Mira Care) | Mira Firmware | 1.7.1.47 | affected |
| Quanovate Tech Inc. (operating as Mira / Mira Care) | Mira Android App | 4.5.15.4 | affected |
Weaknesses
- CWE-290: CWE-290
References
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-223-01
- https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-223-01.json
Feedback
Was this page helpful?
Glad to hear it! Please tell us how we can improve.
Sorry to hear that. Please tell us how we can improve.