CVE-2026-0611
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N
Summary
Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.
Affected Software
| Vendor | Product | Version Range | Status |
|---|---|---|---|
| Spacelabs Healthcare | Sentinel | 10.5.0 < 11.6.0 | affected |
Weaknesses
- CWE-306: Missing Authentication for Critical Function
ADP Enrichment
CISA ADP Vulnrichment
- SSVC:
- Exploitation: poc
- Automatable: yes
- Technical Impact: total
Additional References
References
- https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory-Sentinel-.NET-Remoting-Vulnerability.pdf
- https://spacelabshealthcare.com/products/diagnostic-cardiology/connectivity/sentinel
- https://www.vulncheck.com/advisories/spacelabs-healthcare-sentinel-x-unauthenticated-rce-via-net-remoting
Feedback
Was this page helpful?
Glad to hear it! Please tell us how we can improve.
Sorry to hear that. Please tell us how we can improve.