CVE-2025-4394
6.8
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Summary
Medtronic MyCareLink Patient Monitor uses an unencrypted filesystem on internal storage, which allows an attacker with physical access to read and modify files.
This issue affects MyCareLink Patient Monitor models 24950 and 24952: before June 25, 2025
Affected Software
| Vendor | Product | Version Range | Status |
|---|---|---|---|
| Medtronic | MyCareLink Patient Monitor 24950 | 0 < June 25, 2025 | affected |
| Medtronic | MyCareLink Patient Monitor 24952 | 0 < June 25, 2025 | affected |
Weaknesses
- CWE-312: CWE-312 Cleartext Storage of Sensitive Information
ADP Enrichment
CISA ADP Vulnrichment
- SSVC:
- Exploitation: none
- Automatable: no
- Technical Impact: total
References
- https://www.medtronic.com/en-us/e/product-security/security-bulletins/mycarelink-patient-monitor-vulnerabilities.html
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-205-01