CVE-2025-15653

Summary

Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.

Affected Software

VendorProductVersion RangeStatus
DrägerZeus IE0 < 1.0.5affected
DrägerZeus RS C5000 < 1.0.9affected

Weaknesses

  • CWE-668: CWE-668 Exposure of Resource to Wrong Sphere

ADP Enrichment

CISA ADP Vulnrichment

  • SSVC:
  • Exploitation: none
    • Automatable: no
    • Technical Impact: total

References