CVE-2025-15653
7
CVSS:4.0/AV:P/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N
Summary
Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.
Affected Software
| Vendor | Product | Version Range | Status |
|---|---|---|---|
| Dräger | Zeus IE | 0 < 1.0.5 | affected |
| Dräger | Zeus RS C500 | 0 < 1.0.9 | affected |
Weaknesses
- CWE-668: CWE-668 Exposure of Resource to Wrong Sphere
ADP Enrichment
CISA ADP Vulnrichment
- SSVC:
- Exploitation: none
- Automatable: no
- Technical Impact: total
References
- https://static.draeger.com/security/download/Product-Security-Advisory-25-349-Zeus.pdf
- https://www.vulncheck.com/advisories/dr-ger-zeus-ie-anesthesia-workstation-usb-interface-privilege-escalation
Feedback
Was this page helpful?
Glad to hear it! Please tell us how we can improve.
Sorry to hear that. Please tell us how we can improve.