CVE-2025-14095

Summary

A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software. 

Other related CVE's are CVE-2025-14096 & CVE-2025-14097.

Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.

Required configuration for Exposure:

Physical access to the analyzer is needed.

Temporary work Around:

Only authorized people can physically access the analyzer.

Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution.

Exploit Status: Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication.                                                                                                                                                                                        Note:

CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.

Affected Software

VendorProductVersion RangeStatus
Radiometer Medical ApsABL90 FLEX and ABL90 FLEX PLUS AnalyzersAll application software versions with Windows 7, Windows XP as underlying OSaffected
Radiometer Medical ApsABL90 FLEX and ABL90 FLEX PLUS AnalyzersApplication software versions < 3.5MR11 with Windows 10 as underlying OSaffected
Radiometer Medical ApsAQT90 FLEX AnalyzersAll Application software versions <= 8.13 MR2affected
Radiometer Medical ApsABL800 BASIC and ABL800 FLEX AnalyzersApplication software versions < 6.20 MR2 with Windows 7, Windows XP as underlying OSaffected
Radiometer Medical ApsABL800 BASIC and ABL800 FLEX AnalyzersApplication software versions < 6.20 MR2 with Windows 10 as underlying OSaffected
Radiometer Medical ApsABL9 AnalyzersApplication software versions < 1.5.0affected

Weaknesses

  • CWE-284: CWE-284: Improper Access Control
  • CWE-693: CWE-693 — Protection Mechanism Failure

Workarounds

Only authorized people can physically access the analyzer.

ADP Enrichment

CISA ADP Vulnrichment

  • SSVC:
  • Exploitation: none
    • Automatable: no
    • Technical Impact: partial