CVE-2025-1204
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N
Summary
The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.
Affected Software
| Vendor | Product | Version Range | Status |
|---|---|---|---|
| Contec Health | CMS8000 Patient Monitor | 0 | affected |
Weaknesses
- CWE-912: CWE-912 Hidden Functionality
Workarounds
Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.
If asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.
ADP Enrichment
CISA ADP Vulnrichment
- SSVC:
- Exploitation: poc
- Automatable: no
- Technical Impact: total
References
- https://claroty.com/team82/research/are-contec-cms8000-patient-monitors-infected-with-a-chinese-backdoor-the-reality-is-more-complicated?ref=vault33.org
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01