CVE-2024-9832

Summary

There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.

Affected Software

VendorProductVersion RangeStatus
BaxterLife2000 Ventilation System06.08.00.00 and prioraffected

Weaknesses

  • CWE-307: CWE-307 Improper Restriction of Excessive Authentication Attempts

ADP Enrichment

CISA ADP Vulnrichment

  • SSVC:
  • Exploitation: none
    • Automatable: no
    • Technical Impact: total

References