CVE-2024-48971
9.3
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Summary
The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges.
Affected Software
| Vendor | Product | Version Range | Status |
|---|---|---|---|
| Baxter | Life2000 Ventilation System | 06.08.00.00 and prior | affected |
Weaknesses
- CWE-798: CWE-798 Use of Hard-coded Credentials
ADP Enrichment
CISA ADP Vulnrichment
- SSVC:
- Exploitation: none
- Automatable: no
- Technical Impact: total