CVE-2022-43557
CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:H
Summary
The BD BodyGuard™ infusion pumps specified allow for access through the RS-232 (serial) port interface. If exploited, threat actors with physical access, specialized equipment and knowledge may be able to configure or disable the pump. No electronic protected health information (ePHI), protected health information (PHI) or personally identifiable information (PII) is stored in the pump.
Affected Software
| Vendor | Product | Version Range | Status |
|---|---|---|---|
| Becton, Dickinson and Company (BD) | BodyGuard™ Pump | BD BodyGuard™ | affected |
| Becton, Dickinson and Company (BD) | BodyGuard™ Pump | CME BodyGuard™ 323 (2nd Edition) | affected |
| Becton, Dickinson and Company (BD) | BodyGuard™ Pump | CME BodyGuard™ 323 Color Vision (2nd Edition) | affected |
| Becton, Dickinson and Company (BD) | BodyGuard™ Pump | CME BodyGuard™ 323 Color Vision (3rd Edition) | affected |
| Becton, Dickinson and Company (BD) | BodyGuard™ Pump | CME BodyGuard™ Twins (2nd Edition) | affected |
Weaknesses
- CWE-1299: CWE-1299 Missing Protection Mechanism for Alternate Hardware Interface
Workarounds
BD recommends the following mitigations and compensating controls to reduce risk associated with this vulnerability: • Ensure physical access controls are in place and only authorized end-users have access to BD BodyGuard™ pumps. • Ensure that only BD-approved equipment is connected to the RS-232 interface of the affected pumps. • Ensure that no equipment is connected to the RS-232 interface when the affected pumps are delivering infusions. • Protect connected computer systems with BodyComm™ software with standard security measures.
ADP Enrichment
CVE Program Container
Additional References
CISA ADP Vulnrichment
- SSVC:
- Exploitation: none
- Automatable: no
- Technical Impact: partial
References
Feedback
Was this page helpful?
Glad to hear it! Please tell us how we can improve.
Sorry to hear that. Please tell us how we can improve.