CVE-2022-38453
CVSS:3.1/AV:L/AC:H/PR:H/UI:N/S:U/C:L/I:L/A:N
Summary
Multiple binary application files on the CMS8000 device are compiled with 'not stripped' and 'debug_info' compilation settings. These compiler settings greatly decrease the level of effort for a threat actor to reverse engineer sensitive code and identify additional vulnerabilities.
Affected Software
| Vendor | Product | Version Range | Status |
|---|---|---|---|
| Contec Health | CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor | All | affected |
Weaknesses
- CWE-489: CWE-489 Leftover Debug Code
Workarounds
Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.
The following mitigations could assist in reducing the risk for exploitation of vulnerabilities:
Disabling UART functionality at the CPU level Enforcing unique device authentication before granting access to the terminal / bootloader Where possible, enforcing secure boot. Tamper stickers on the device casing to indicate when a device has been opened
ADP Enrichment
CVE Program Container
Additional References
CISA ADP Vulnrichment
- SSVC:
- Exploitation: none
- Automatable: no
- Technical Impact: partial
References
Feedback
Was this page helpful?
Glad to hear it! Please tell us how we can improve.
Sorry to hear that. Please tell us how we can improve.