CVE-2021-33885
10
CVSS:3.1/AC:L/AV:N/A:N/C:H/I:H/PR:N/S:C/UI:N
Summary
An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This results in full system command access and execution because of the lack of cryptographic signatures on critical data sets.
Affected Software
| Vendor | Product | Version Range | Status |
|---|---|---|---|
| n/a | n/a | n/a | affected |
Weaknesses
- n/a
ADP Enrichment
CVE Program Container
Additional References
- https://www.bbraunusa.com/en.htm
- https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/
References
- https://www.bbraunusa.com/en.htm
- https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/
Feedback
Was this page helpful?
Glad to hear it! Please tell us how we can improve.
Sorry to hear that. Please tell us how we can improve.