CVE-2021-26262
CVSS:4.0/AV:L/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N
Summary
Philips MRI 1.5T and MRI 3T Version 5.3 through 5.8.1 does not restrict or incorrectly restricts access to a resource from an unauthorized actor.
Affected Software
| Vendor | Product | Version Range | Status |
|---|---|---|---|
| Philips | MRI 1.5T | 5.3 <= 5.8.1 | affected |
| Philips | MRI 3T | 5.3 <= 5.8.1 | affected |
Weaknesses
- CWE-286: CWE-286 Incorrect User Management
Workarounds
Philips released a software upgrade version 5.8.2 to remediate these vulnerabilities and can be referenced by FCO78100619.
As an interim mitigation to these vulnerabilities, Philips recommends the following:
Users should operate all Philips deployed and supported products within Philips authorized specifications, including physical and logical controls. Only allowed personnel are permitted in the vicinity of the product. Refer to the Philips instructions for use (IFU) available on InCenter https://incenter.medical.philips.com.
Users with questions about their specific MRI product should contact a Philips service support team or regional service support. Philips contact information is available at the Philips customer service solutions website http://philips.com/productsecurity
or by calling 1-800-722-9377.
For more information regarding these vulnerabilities, see the Philips product security advisory website http://philips.com/productsecurity.
Users can also visit the Philips product security website for the latest security information for Philips products.
ADP Enrichment
CVE Program Container
Additional References
- https://us-cert.cisa.gov/ics/advisories/icsma-21-313-01
- https://www.usa.philips.com/healthcare/about/customer-support/product-security
References
- https://us-cert.cisa.gov/ics/advisories/icsma-21-313-01
- https://www.usa.philips.com/healthcare/about/customer-support/product-security
Feedback
Was this page helpful?
Glad to hear it! Please tell us how we can improve.
Sorry to hear that. Please tell us how we can improve.