CVE-2020-16222

Summary

In Patient Information Center iX (PICiX) Version B.02, C.02, C.03, and PerformanceBridge Focal Point Version A.01, when an actor claims to have a given identity, the software does not prove or insufficiently proves the claim is correct.

Affected Software

VendorProductVersion RangeStatus
PhilipsPatient Information Center iX (PICiX)B.02affected
PhilipsPatient Information Center iX (PICiX)C.02affected
PhilipsPatient Information Center iX (PICiX)C.03affected
PhilipsPerformanceBridge Focal PointA.01affected

Weaknesses

  • CWE-287: CWE-287 Improper Authentication

Workarounds

As a mitigation to these vulnerabilities, Philips recommends the following:

  • The Philips patient monitoring network is required to be physically or logically isolated from the hospital local area network (LAN). Philips recommends using a firewall or routers that can implement access control lists restricting access in and out of the patient monitoring network for only necessary ports and IP addresses. Refer to the Philips Patient Monitoring System Security for Clinical Networks guide for additional information on InCenter https://incenter.medical.philips.com/ .

  • By default, the simple certificate enrollment protocol (SCEP) service is not running. When needed, the service is configured to run based on the duration or the number of certificates to be assigned. One certificate is default, but if a certificate is not issued, the service will continue to run. Limit exposure by ensuring the SCEP service is not running unless it is actively being used to enroll new devices.

  • When enrolling new devices using SCEP, enter a unique challenge password of 8-12 unpredictable and randomized digits.

  • Implement physical security controls to prevent unauthorized login attempts on the PIC iX application. Servers should be kept in controlled locked data centers. Access to equipment at nurses’ stations should be controlled and monitored.

  • Only grant remote access to PIC iX servers on a must-have basis.

  • Grant login privileges to the bedside monitor and PIC iX application on a role-based, least-privilege basis, and only to trusted users.

Users with questions regarding their specific Philips Patient Information Center (PIC iX) and/or IntelliVue patient monitor installations and new release eligibility should contact their local Philips service support team, or regional service support https://www.usa.philips.com/healthcare/solutions/customer-service-solutions , or call 1-800-722-9377.

Please see the Philips product security website https://www.philips.com/productsecurity for the Philips advisory and the latest security information for Philips products.