CVE-2019-25723
6.3
CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L
Summary
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
Affected Software
| Vendor | Product | Version Range | Status |
|---|---|---|---|
| Dräger | Perseus A500 | 2.00 | affected |
| Dräger | Perseus A500 | 2.01 | affected |
| Dräger | Perseus A500 | 2.02 | affected |
| Dräger | Perseus A500 | 2.03 | unaffected |
Weaknesses
- CWE-1286: CWE-1286 Improper Validation of Syntactic Correctness of Input
ADP Enrichment
CISA ADP Vulnrichment
- SSVC:
- Exploitation: none
- Automatable: no
- Technical Impact: partial
References
- https://static.draeger.com/security/download/2020-06-PSA-20-169-01-Security-Advisory-Perseus-A500-SW20n.pdf
- https://www.vulncheck.com/advisories/dr-ger-perseus-a500-dos-via-medibus-interface