CVE-2019-25721
7.1
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N
Summary
Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.
Affected Software
| Vendor | Product | Version Range | Status |
|---|---|---|---|
| Dräger | Infinity M300 | 0 <= VG2.3.1 | affected |
| Dräger | Infinity M300 | VG2.3.2 | unaffected |
Weaknesses
- CWE-400: CWE-400 Uncontrolled Resource Consumption
ADP Enrichment
CISA ADP Vulnrichment
- SSVC:
- Exploitation: none
- Automatable: no
- Technical Impact: partial
References
- https://static.draeger.com/security/download/2019-09-13-Cybersecurity-M300-VGxx-Security-advisory-update_v1-3.pdf
- https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-network-based-denial-of-service