CVE-2019-25719
CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N
Summary
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
Affected Software
| Vendor | Product | Version Range | Status |
|---|---|---|---|
| Dräger | Infinity Acute Care System | 0 <= VG4.1.1 | affected |
| Dräger | Infinity Acute Care System | 0 <= VG4.0.3 | affected |
| Dräger | Infinity Acute Care System | VG4.2 | unaffected |
| Dräger | Standalone Infinity M540 patient monitor | 0 <= VG4.1.1 | affected |
| Dräger | Standalone Infinity M540 patient monitor | VG4.2 | unaffected |
Weaknesses
- CWE-924: CWE-924 Improper Enforcement of Message Integrity During Transmission in a Communication Channel
ADP Enrichment
CISA ADP Vulnrichment
- SSVC:
- Exploitation: none
- Automatable: yes
- Technical Impact: partial
References
- https://static.draeger.com/security/download/PSA-19-255-02_Product-Security-Advisory-IACS-VG4.1.pdf
- https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofing-and-dos-via-network-message-handling