CVE-2015-3954

Summary

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Affected Software

VendorProductVersion RangeStatus
HospiraPlum A+ Infusion System<= 13.4affected
HospiraPlum A+3 Infusion System<= 13.6affected
HospiraSymbiq Infusion System<= 3.13affected

Weaknesses

  • CWE-285: Improper authorization CWE-285

ADP Enrichment

CVE Program Container

Additional References

References